How can I estimate the minimum current measurement of a current probe using an oscilloscope?
September 27, 2021
Here is a quick method for estimating the lowest current measurement that is possible with an oscilloscope and current probe. This is especially useful when the datasheet for a probe may not show the minimum current that can be measured or it may be missing the noise floor value. For this note, we will be … Continued
Do your instruments support UDP communications over LAN?
August 5, 2021
Not at this time. Many instruments support communication via sockets (port 5025) and telnet (port 5024) but the port numbers cannot be changed. If you use VISA, you can implement LAN via TCP/IP, which does not need a port number. Here is a listing of current SIGLENT instruments with Socket and Telnet Support.
How to check available internal memory?
July 29, 2021
Simply access the Save/Recall menu for the instrument in question and select Local. The menu will show used/total amounts. Here, the SDG6X has approximately 83 MB and the SDG1X has 85 MB of memory:
Bode plot memory depths
July 19, 2021
Here is a table showing the various memory depths used in the Bode Plot feature Product series Max memory depth SDS1xx4X-E 140k SDS2000X-E 140k SDS2000X Plus 100k SDS5000X 125k
Why is the Bode Plot start frequency minimum 10 Hz?
June 23, 2021
The minimum start frequency is set to 10 Hz because of the extended acquisition time required to collect enough data to ensure a useful level of measurement accuracy. If the low limit was 1 Hz, the scope would need to test 10 cycles (to maintain specified accuracy), so the time base will be 10 s/div. … Continued
SDS2000X Plus and SDS5000X Default Logic Levels
May 17, 2021
The SDS2000X Plus and SDS5000X series oscilloscopes have default threshold and trigger values for the MSO/Digital channels listed below: TTL = 1.5V CMOS = 1.65V LVCMOS33 = 1.65V LVCMOS25 = 1.25V
May 7, 2021
Per OSHA, a Nationally Recognized Testing Laboratory (NRTL) is an independent third-part laboratory recognized by the United States Occupational Safety and Health Administration (OSHA) to test and certify products to applicable product safety standards. Examples of NRTLs include UL, CSA, and TUV. Many SIGLENT products are TUV certified, which also covers UL and CSA requirements. To … Continued
Are SIGLENT products certified by a Nationally Recognized Test Laboratories (NRTL)?
May 7, 2021
Per OSHA, a Nationally Recognized Testing Laboratory (NRTL) is an independent third-part laboratory recognized by the United States Occupational Safety and Health Administration (OSHA) to test and certify products to applicable product safety standards. Examples of NRTLs include UL, CSA, and TUV. Many SIGLENT products are TUV certified, which also covers UL and CSA requirements. To … Continued
SAG1021/SAG1021I output current limit
April 6, 2021
30 mA is the theoretical maximum for SAG1021/SAG1021I. I = V/R = 1.5V(max) / 50 Ohm = 30 mA
SDS2000X Plus SDS5000X Series USB/LAN specifications
March 30, 2021
USB device: USB 2.0, HS (high speed) 480Mbps LAN:100Mbps